Kit related FAQ - Why are the retention times shifted?

If the retention times are unstable across a plate or shifting over time:

  • Make sure the mobile phases are purged and there are no bubbles in the system.
  • Make sure no air is drawn or injected by the autosampler.
  • Make sure the entire LC-MS system is well conditioned and the column is sufficiently equilibrated.

Otherwise, check the following parts with your instrument engineer:

  • Degasser
  • Mixing chamber (needs to be replaced?)
  • Intake/outlet valve at pump head
  • Seals at pump head
  • Replace the analytical column (it may have reached the end of its lifespan)